Techniques

Gas Chromatography-Mass Spectrometry

GC-MS

GC-MS system coupled with:

  • KAUST closed batch reactor with top illumination window
  • Nevolab closed batch glass reactor with inside illumination
  • Light sources: LEDs (365, 405, 450, 525 nm), AAA solar simulator
Scheme of our CO₂ conversion setup with photoreactor and GC-MS
Our home-built CO₂ conversion setup

The closed batch system consists of an isolated stainless-steel or glass reactor connected to a gas chromatographer/mass-spectrometer via a transfer line. Both, automatic injection via transfer line and manual injection, are possible.

We use GC-MS Shimadzu QP2020-NX with electron ionisation (EI), which is equipped with a U-Bond column (30 m, ID 0.32 mm) for CO₂ separation and a Molsieve 5A column (30 m, ID 0.32 mm) for CO separation without N₂ interference. The system allows us to separate and quantify CO₂, C1–C6 hydrocarbons, C1–C4 oxygenated hydrocarbons (liquids and gases), as well as permanent gases. Helium is used as a carrier gas to improve the sensitivity for hydrocarbons.

KAUST closed batch reactor with quartz window, O-ring and PTFE sample holder
KAUST closed batch reactor

Before the CO₂ conversion reaction, the reactor is typically evacuated and purged with N₂ at least 6 times to remove air. Then the reactor is filled with CO₂ gas to a chosen pressure. During the reaction, the photocatalyst suspension is stirred and illuminated with one of the LEDs or a solar simulator. At the end of the reaction, the transfer line is evacuated to 0.1 bar before the analysis. The gas mixture is injected into the GC sample loop via the pressure drop between the reaction chamber and the transfer line. The ability to evacuate the transfer line and the sample loop prior to the analysis eliminates the possibility of contamination with air and provides a higher sensitivity for targeted analytes.

Importantly, the origin of generated CO₂ conversion products can also be easily confirmed with ¹³C-isotope labelling using our GC-MS system with ¹³C isotope labelled CO₂ gas (Sigma Aldrich, 99 atom % ¹³C, < 3 atom % ¹⁸O).

Illuminated photoreactor during a measurement

Want to build a reactor yourself? The CAD drawings are available here: Make your own photocatalysis reaction chamber